MT-3109 AC/DC Digital Clamp Multimeter
C. Resistance
Range Resolution Accuracy
Overload
protection
400.0 100m +(1.2%+5)
4.000k 1
40.00k 10
400.0k 10
+(1 %+5)
4.000M 1k +(1.2%+5)
40.00M 10k +(1.5%+5)
600Vp
D. Diode Test
Range Resolution Accuracy Overload
protection
1mV
Display forward
voltage drop
nearest value
600Vp
Remark: Open circuit voltage approximate 1.48V.
E. Continuity Test
Range resolution accuracy Overload
protection
100m Around50
The buzzer beeps
600Vp
Remark:
Open circuit voltage approximate 0.45V.
The buzzer may or may not beeps when the resistance of a
circuit under test is between 50~1OO
The buzzer will not beep when the resistance of a circuit under
test is > 1OOQ.